149

High Voltage Meter Series

Applications:

  • 4½ digit LCD display
  • High input impedance
  • Positive and negative reading without
    switching
  • Voltage coefficient compensated
  • Accuracy better than ±0.05%
  • Precision output for recorder or oscilloscope
  • 40kV hand probe supplied with 149-01 & -03
  • Rugged construction

Electrical Specification

  149-01 & 149-03 only

Voltage ranges

0 to 2kV dc, 20kV dc & 40kV dc.

Resolution 0.1V on 2kV range, 1V on 20kV range
and 10V on 40kV range.
Accuracy ±0.1% ±2 digits on the 2kV, 20kV & 40kV ranges
HV input Suppied with plug & 40kV hand held
probe (with 2m of cable,see drawings).
  149-02 & 149-04 only
Voltage ranges 0 to 2kV dc, 20kV dc & 60kV dc.
Resolution 0.1V on 2kV range, 1V on 20kV range
and 10V on 60kV range.
Accuracy ±0.05% ±2 digits on the 2kV, 20kV & 60kV ranges
HV input Suppied with plug & 1m of screened
60kV cable assembly.
  149-01 and 149-02 only
Input resistance 10GΩ on all ranges
  149-03 and 149-04 only
Input resistance 100GΩ on all ranges
  149 (all in the series)
Stability Better than 0.3% over 1 year.
Polarity Positive or negative, auto reading.
Protection Circuits protected against damage by
external arcing, flashover etc
Recorder output 10mV per kV (Source/sink capability up
to 100μA.)
Accuracy ±0.3% at 30kV (±0.5% at the
extremes).

Environmental Specification

Operating temperature

  0 to +50 deg.C

Storage temperature

  -20 to +65 deg.C

Humidity (RH)

<31°C
>30°C
80% maximum.decrease linearly to 50% at 40°C (non-condensing)
For indoor use only.

Download PDF

Download PDF:

Data Sheet