power supplies, scientific instruments scientific instruments
High Voltage Power Supplies
power supplies, photomultiplier tubes
PRODUCTS | COMPANY PROFILE | FAQ's | CAREERS | NEWS & EVENTS | SUPPORT | CONTACT US | HOME
 

 

Precision high voltage modules

 

LT Series Data Sheet
LT001, LT2.5, LT005, LT010,
LT015, LT020, LT030, LT050

PRECISION HIGH VOLTAGE MODULES

Applications : (3 Watt max)

- Mass spectrometers
- Electron microscopes
- Nuclear instruments
- Photomultiplier tubes
- Microchannel plates

Precision high voltage power supply modules

[Click image to enlarge]


  • 1kV, 2.5kV, 5kV, 10kV, 15kV, 20kV, 30kV, 50kV
  • Highest Stability, Lowest Ripple, Lowest Drift
  • Externally programmable or Internal control (option)
  • Short circuit and flashover proof
  • 24 hour burn in
  • Low radiated magnetic field
  • Positive, negative & remote reversible versions


The LT Series builds on the new standard for high voltage precision and stability set by the LS 1 W family. By optimising the Low Stress™ technologies pioneered in the LS family, the LT Series provides 3W o/p capability, but still has less than 2ppm ripple, 10ppm/°C temperature coefficient, and 75ppm drift/hr throughout the range, while retaining the same LS case dimensions. Each of the units includes a differential control input.

To take full advantage of the performance available from the LT Range, they are initially only being made available to our OEM customers as application specific, custom optimised, modules.

All units are short circuit proof and use proven techniques to drive a high frequency oscillator and ferrite high voltage step-up transformer. These advanced power supplies build on Applied Kilovolts’ legendary reputation for reliability, being constructed from conservatively rated components with their reliability further enhanced by information gained over many years of field operation.

Available as positive or negative up to 50kV, and electrically reversible at voltages up to 30kV.

We manufacture a large number of customized OEM versions and would be pleased to discuss your application with you.


Electrical Specification
Positive & Negative Polarity Units

UNIT TYPE

POLARITY

OUTPUT

RIPPLE AT FULL LOAD

SIZE (mm)

  LT001PIP010

POSITIVE

10 volts to 1kV at 3mA

2mV peak to peak

156 X 127 X 64

  LT001NIP010

NEGATIVE

  LT2.5PIP010

POSITIVE

10 volts to 2.5kV at 1.2mA

5mV peak to peak

156 X 127 X 64

  LT2.5NIP010

NEGATIVE

  LT005PIP010

POSITIVE

10volts to 5kV at 600uA

10mV peak to peak

156 X 127 X 64

  LT005NIP010

NEGATIVE

  LT010PIP010

POSITIVE

20 volts to 10kV at 300uA

20mV peak to peak

156 X 127 X 64

  LT010NIP010

NEGATIVE

  LT015PIP010

POSITIVE

30 volts to 15kV at 200uA

30mV peak to peak

207 X 148 X 74

  LT015NIP010

NEGATIVE

  LT020PIP010

POSITIVE

50 volts to 20kV at 150uA

40mV peak to peak

207 X 148 X 74

  LT020NIP010

NEGATIVE

  LT030PIP010

POSITIVE

100 volts to 30kV at 100uA

60mV peak to peak

207 X 148 X 74

  LT030NIP010

NEGATIVE

  LT050PIP010

POSITIVE

200 volts to 50kV at 60uA

100mV peak to peak

333 X 204 X 114

  LT050NIP010

NEGATIVE


Electrical Specification
REVERSIBLE POLARITY UNITS

UNIT TYPE

POLARITY

OUTPUT

RIPPLE AT FULL LOAD

SIZE (mm)

  LT001RIP010

REVERSIBLE

±10V to ±1kV at 3mA

2mV peak to peak

163 X 152 X 71.5

  LT2.5RIP010

REVERSIBLE

±10V to ±2.5kV at 1.2mA

5mV peak to peak

163 X 152 X 71.5

  LT005RIP010

REVERSIBLE

±10V to ±5kV at 600uA

10mV peak to peak

163 X 152 X 71.5

  LT010RIP010

REVERSIBLE

±20V to ±10kV at 300uA

20mV peak to peak

163 X 152 X 71.5

  LT015RIP010

REVERSIBLE

±30V to ±15kV at 200uA

30mV peak to peak

233 X 204 X 114

  LT020RIP010

REVERSIBLE

±50V to ±20kV at 150uA

40mV peak to peak

233 X 204 X 114

  LT030RIP010

REVERSIBLE

±100V to ±30kV at 100uA

60mV peak to peak

333 X 204 X 114

Please see the HPZ data sheet for continuously variable zero crossing power supplies.

 

INPUT VOLTAGE:

+24 volt d.c. ±1V at <500mA, for Unipolar, and at <600mA for Reversibles. Negative input terminal common to HV earth return.

OUTPUT VOLTAGE:

See tables above.

LINE REGULATION:

Less than 4 ppm of maximum voltage for input changes of 1 volt.

LOAD REGULATION:

Less than 10ppm for load changes from 10% to 100%.
(Measured at maximum voltage)

OVERLOAD PROTECTION:

Flashover and short circuit proof. Re-entrant current limit.

TEMPERATURE CO-EFFICIENT:

<10ppm/°C from 20°C to 50°C

DRIFT:

<75ppm per hour (<0.005%)
<150ppm per 8 hours (<0.01%)
Measured at constant input voltage, load current & ambient temperature after 0.5 hour warm up.

CONTROL:

10V analogue signal. (0 to +10V gives zero to maximum output, tolerance +2%, -½%).
Input impedance > 100Kohm.
EXTERNAL potentiometer = Option Code PP
INTERNAL potentiometer = Option Code FP

READOUT:

Voltage monitor : 0 to +10V represents zero to maximum output, tolerance ±2% (Zout=10k).
Precision Current Monitor :
0 to 10V represents zero to 100% o/p current, ±2%, Offset ±0.1% of FS (Zout=10k)

OPERATING TEMPERATURE:

0°C to +50°C

STORAGE TEMPERATURE:

-35°C to +85°C

R.F.I.:

Steel case for low radiated magnetic field.

 

Mechanical Specification

FIXINGS:

Blind Fasteners into the base.

OUTPUT:

By 1 metre screened cable, or optional integrated high voltage connector
N.B. Reducing the cable length may increase the ripple voltage.

INPUT CONNECTOR:

15 way D-type


E.g Order Code:
LT010NIP010 = negative 10kV unit
LT005PPP010 = positive 5kV unit for control by External Potentiometer.

Download PDF

Download PDF:

Data Sheet | LT Installation Notes | LTR Installation Notes | Outline Drawing |


Top
| LS Detailed Photo | LT Safety Instructions | LTR Safety Instructions

 

power supplies
power supply manufacturer
Power Supplies | Units by o/p Voltage | Company Profile | FAQ's | Contact | Careers | Site Map | Home