| LT Series Data Sheet
LT001, LT2.5, LT005, LT010,
LT015, LT020, LT030, LT050
PRECISION HIGH VOLTAGE MODULES
Applications : (3 Watt max)
- Mass spectrometers
- Electron microscopes
- Nuclear instruments
- Photomultiplier tubes
- Microchannel plates
|

[Click image to enlarge] |
- 1kV, 2.5kV, 5kV, 10kV, 15kV, 20kV, 30kV, 50kV
- Highest Stability, Lowest Ripple, Lowest Drift
- Externally programmable or Internal control (option)
|
- Short circuit and flashover proof
- 24 hour burn in
- Low radiated magnetic field
- Positive, negative & remote reversible versions
|
|
|
The LT Series builds on the new standard for high voltage precision
and stability set by the LS 1 W family.
By optimising the Low Stress™ technologies pioneered in the
LS family, the LT Series provides 3W o/p capability, but still has
less than 2ppm ripple, 10ppm/°C temperature coefficient, and
75ppm drift/hr throughout the range, while retaining the same LS
case dimensions. Each of the units includes a differential control
input.
To take full advantage of the performance available from the
LT Range, they are initially only being made available to our OEM
customers as application specific, custom optimised, modules.
All units are short circuit proof and use proven techniques to
drive a high frequency oscillator and ferrite high voltage step-up
transformer. These advanced power supplies build on Applied Kilovolts’
legendary reputation for reliability, being constructed from conservatively
rated components with their reliability further enhanced by information
gained over many years of field operation.
Available as positive or negative up to 50kV, and electrically
reversible at voltages up to 30kV.
We manufacture a large number of customized OEM versions and would
be pleased to discuss your application with you.
|
Electrical Specification
Positive & Negative Polarity Units |
| UNIT TYPE |
POLARITY |
OUTPUT |
RIPPLE AT FULL LOAD |
SIZE (mm) |
LT001PIP010 |
POSITIVE |
10 volts to
1kV at 3mA |
2mV peak to
peak |
156 X 127 X
64 |
LT001NIP010 |
NEGATIVE |
LT2.5PIP010 |
POSITIVE |
10 volts to
2.5kV at 1.2mA |
5mV peak to
peak |
156 X 127 X
64 |
LT2.5NIP010 |
NEGATIVE |
LT005PIP010 |
POSITIVE |
10volts to 5kV
at 600uA |
10mV peak to
peak |
156 X 127 X
64 |
LT005NIP010 |
NEGATIVE |
LT010PIP010 |
POSITIVE |
20 volts to
10kV at 300uA |
20mV peak to
peak |
156 X 127 X
64 |
LT010NIP010 |
NEGATIVE |
LT015PIP010 |
POSITIVE |
30 volts
to 15kV at 200uA |
30mV peak to
peak |
207 X 148 X
74 |
LT015NIP010 |
NEGATIVE |
LT020PIP010 |
POSITIVE |
50 volts to
20kV at 150uA |
40mV peak to
peak |
207 X 148 X
74 |
LT020NIP010 |
NEGATIVE |
LT030PIP010 |
POSITIVE |
100 volts
to 30kV at 100uA |
60mV peak to
peak |
207 X 148 X
74 |
LT030NIP010 |
NEGATIVE |
LT050PIP010 |
POSITIVE |
200 volts to
50kV at 60uA |
100mV peak to
peak |
333 X 204 X
114 |
LT050NIP010 |
NEGATIVE |
|
Electrical Specification
REVERSIBLE POLARITY UNITS |
| UNIT TYPE |
POLARITY |
OUTPUT |
RIPPLE AT FULL LOAD |
SIZE (mm) |
LT001RIP010 |
REVERSIBLE |
±10V to ±1kV
at 3mA |
2mV peak to peak |
163 X 152 X 71.5 |
LT2.5RIP010 |
REVERSIBLE |
±10V to ±2.5kV
at 1.2mA |
5mV peak to peak |
163 X 152 X 71.5 |
LT005RIP010 |
REVERSIBLE |
±10V to ±5kV
at 600uA |
10mV peak to peak |
163 X 152 X 71.5 |
LT010RIP010 |
REVERSIBLE |
±20V to ±10kV
at 300uA |
20mV peak to peak |
163 X 152 X 71.5 |
LT015RIP010 |
REVERSIBLE |
±30V to ±15kV
at 200uA |
30mV peak to peak |
233 X 204 X 114 |
LT020RIP010 |
REVERSIBLE |
±50V to ±20kV
at 150uA |
40mV peak to peak |
233 X 204 X 114 |
LT030RIP010 |
REVERSIBLE |
±100V to ±30kV
at 100uA |
60mV peak to peak |
333 X 204 X 114 |
| Please see the HPZ
data sheet for continuously variable zero crossing power supplies.
|
|
| INPUT
VOLTAGE: |
+24 volt d.c. ±1V
at <500mA, for Unipolar, and at <600mA for Reversibles.
Negative input terminal common to HV earth return. |
| OUTPUT VOLTAGE: |
See tables above. |
| LINE
REGULATION: |
Less than 4 ppm of maximum voltage for
input changes of 1 volt. |
| LOAD
REGULATION: |
Less than 10ppm for load changes from
10% to 100%.
(Measured at maximum voltage) |
| OVERLOAD PROTECTION: |
Flashover and short circuit proof. Re-entrant
current limit. |
| TEMPERATURE CO-EFFICIENT: |
<10ppm/°C from 20°C to 50°C |
| DRIFT: |
<75ppm per hour (<0.005%)
<150ppm per 8 hours (<0.01%)
Measured at constant input voltage, load current & ambient
temperature after 0.5 hour warm up. |
| CONTROL: |
10V analogue signal. (0 to
+10V gives zero to maximum output, tolerance +2%, -½%).
Input impedance > 100Kohm.
EXTERNAL potentiometer = Option Code PP
INTERNAL potentiometer = Option Code FP |
| READOUT:
|
Voltage monitor :
0 to +10V represents zero to maximum output, tolerance ±2%
(Zout=10k).
Precision Current Monitor : 0 to 10V represents zero to
100% o/p current, ±2%, Offset ±0.1% of FS (Zout=10k)
|
| OPERATING
TEMPERATURE: |
0°C to +50°C |
| STORAGE
TEMPERATURE: |
-35°C to +85°C |
| R.F.I.: |
Steel case for low radiated
magnetic field. |
|
Mechanical Specification |
| FIXINGS: |
Blind Fasteners into the
base. |
| OUTPUT: |
By 1 metre screened cable,
or optional integrated high voltage connector
N.B. Reducing the cable length may increase the ripple voltage. |
INPUT CONNECTOR: |
15 way D-type |
|
E.g Order Code:
LT010NIP010 = negative 10kV unit
LT005PPP010 = positive 5kV unit for control by External Potentiometer.
Top | LS Detailed Photo | LT