Input voltages:
|
+24 V ±10% at 1A.
|
|
Bias Output voltage:
|
See table above. Joined within the unit to Detector -ve
|
|
Bias Control voltage:
|
0V to +10 V gives 0V to max. kV ±2%.
nb: Internal manual potentiometer control not fitted.
|
|
Load regulation:
|
Less than 100 ppm for zero to max. load.
|
|
Line regulation:
|
Less than 100 ppm for 1V change in I/P volts.
|
|
V monitor:
|
0 to ±10V represents 0 to ±max o/p voltage, ±2% output impedance 10 kohm.
|
|
I monitor:
|
0 to ±10V represents 0 to ±max o/p current, ±5% output impedance 10 kohm.
|
|
Speed of reversing:
|
< 500 mS.
|
|
Temperature Co-efficient:
|
<25 ppm per °C. (<10ppm per °C available shortly)
|
|
Detector Control Voltage:
|
0V to +10 V gives 0V to 3 kV ±5%.
|
|
Minimum Detector O/P V:
|
500V
|
|
Detector Vmon:
|
0 to 10 V represents 0 to 3kV o/p voltage, output impedance 10 kohm.
|
|
Detector Load regulation:
|
< 0.1% for zero to max. load.
|
|
Detector Line regulation:
|
< 0.1% for 1V change in I/P volts.
|
|
Detector Temp-co:
|
<300 ppm per °C.
|
|
Operating temperature:
|
0 °C to +45 °C.
|